Light Hydrocarbon Chromatograph With Rotating Disk Autosampler(LH-QTSEY VI)
1)Product principle
Light hydrocarbon chromatograph is a special instrument used in geological logging. It can analyze light hydrocarbon of natural gas, crude oil and rock, and get concentration or relative content of 103 monomer hydrocarbons before C9. It's mainly used for the evaluation of the oil-water layer, the reservoirs water flooded degree and washing degree in later period of oilfield development.
LH-QTSEY VI is a new generation of high-performance intelligent network light hydrocarbon chromatograph. Compared with the traditional instrument, it has five advantages: the fast cooling oven, high detection sensitivity, dedicated workstation software, computer control and networked remote supervision.
Rotating disk autosampler is specially developed for light hydrocarbon analysis, which can improve the automation degree of the process. It uses the sophisticated electronic control unit, to replace the traditional manual mode, which implements multiple samples analysis automatically and continuously, eliminates the human error, improve the working efficiency, analysis accuracy and repeatability.
LH-QTSEY VI light hydrocarbon chromatograph achieved connection between the instrument and autosampler, which can automatic quantitative sample analysis. The instrument provides more comprehensive information for oil and gas layers' estimation, and it is the essential equipment for finding oil-gas layer and evaluating reservoirs in the process of oil exploration logging.
2)Main Technical Parameters
◎Range of temperature control:
Column box: 8℃-400℃ (above environment temperature)
Vaporizing chamber, detector: 20℃-380℃ (above environment temperature)
Switch valve: 20℃-160℃ (above environment temperature)
◎Temperature stability of column box (70℃): <0.5%
◎Temperature program of column box: 8 stages, 0-30℃/min (minute increment 0.1℃/min)
◎Column box temperature programmed repeatability: ≤1% (50℃-250℃, 10℃/min)
◎Cooling time of column box: ≤6min (environment temperature≤27℃)
◎Detection limit of FID: Mt≤ 5×10-12g/s (packed column SE-30, 135℃, nC16 Standard solution)
◎Baseline stability of FID:
Base line noise <10×10-14A,
Base line wander <6×10-13A/30min
◎Power of autosampler: 450W
◎Power of chromatograph: 1500W
◎Maximum number of consecutive samples analyzed: 12
3)Technical Advantages
★ Double gas path system is used in light hydrocarbon chromatograph. Hydrogen as carrier gas makes peaks fast and complete.
★ With feedback control function, the stability of the instrument is improved.
★ Through the network function, the running status of the instrument, sample analysis result and so on can be real-time transferred.
★ Has the function of self-diagnosis of operational failure, failure alarm and cutting off heating power automatically.
★ Has the function of drilling fluid, debris mixed sample pretreatment.
★ Quantitative, automatic, continuous sampling analysis.
★ Autosampler can be used with AGILENT 7890 and 7820, achieve automatic and continuous analysis, to meet the different needs of all users.